Continuous glucose monitoring and use of alternative markers to assess glycemia in chronic kidney disease

Download (562.62 kB)
figure
posted on 2020-08-11, 17:41 authored by Leila R. Zelnick, Zona O. Batacchi, Ashveena Dighe, Randie R. Little, Dace L. Trence, Irl B. Hirsch, Ian H. de Boer
Objective: In chronic kidney disease (CKD), glycated albumin and fructosamine have been postulated to be better biomarkers of glycemic control than HbA1c. We evaluated accuracy, variability, and covariate bias of three biomarkers (HbA1c, glycated albumin, and fructosamine) compared with continuous glucose monitoring (CGM)-derived measurement of glycemia across estimated glomerular filtration rate (eGFR) in type 2 diabetes.

Research Design and Methods: Prospective cohort study of 104 participants with type 2 diabetes, 80 with eGFR <60 mL/min/1.73m2 (not treated with dialysis) and 24 frequency-matched controls with eGFR ≥60 mL/min/1.73m2. Participants wore a blinded CGM for 2 six-day periods separated by two weeks, with blood and urine collected at the end of each CGM period. HbA1c, glycated albumin and fructosamine were measured by HPLC, enzymatic, and colorimetric nitroblue tetrozolium methods, respectively.

Results: Within-person biomarker values were strongly correlated between the two CGM periods (r=0.92-0.95), although no marker fully captured the within-person variability of mean CGM glucose. All markers were similarly correlated with mean CGM glucose (r = 0.71-77). Compared with mean CGM glucose, glycated albumin and fructosamine were significantly biased by age, BMI, serum iron concentration, transferrin saturation, and albuminuria; HbA1c was under-estimated in those with albuminuria.

Conclusions: Glycated albumin and fructosamine were not less variable than HbA1c at a given mean CGM glucose level, with several additional sources of bias. These results support measuring HbA1c to monitor trends in glycemia among patients with eGFR <60 mL/min/1.73m2. Direct measurements of glucose are necessary to capture short-term variability.

Funding

The Continuous Glucose Monitoring to Assess Glycemia in CKD study was primarily supported by American Diabetes Association grant #4-15-CKD-20. Additional funding came from grants R01DK088762, R01087726, and T32DK007247 from the National Institute of Diabetes and Digestive and Kidney Diseases, a grant from Puget Sound Veterans Affairs Health Care System, and an unrestricted grant from the Northwest Kidney Centers. Continuous glucose monitoring equipment and supplies were donated by Medtronic, and self-monitored blood glucose equipment and supplies were donated by Abbott. Study sponsors had no role in designing the study, collecting study data, or analyzing or presenting study results.